280
34980A User’s Guide
10
64-Bit Digital I/O Module with Memory and Counter
Set the number of samples to collect.
The
SENSe:DIGital:MEMory:SAMPle:COUNt
command sets the number of samples to capture. If you set the number of
counts to infinite (0 = default), the bank will capture data until a STOP is
received. Older samples are overwritten if memory gets full. Allowed
sample counts depend upon the channel width as follows:
•
BYTE (8- bit) 1 to 65535
•
WORD (16- bit) 1 to 65535
•
LWORd (32- bit) 1 to 32767
Start the capture.
The
SENSe:DIGital:MEMory:STARt
command sets the
channel to begin the data capture. The capture begins when the handshake
occurs.
Check the status of the transfer.
You can use the
SENSe:DIGital:MEMory:DATA:POINts?
query to return the number of
samples currently in memory.
Retrieve the captured data.
Set the desired memory retrieval format using
the
SENSe:DIGital:DATA:FORMat
command. You can set the memory to be
read as either
LIST
or
BLOCk
. The
LIST
parameter (default) returns the
data as comma separated ASCII values.
BLOCk
returns the data in
IEEE- 488 block format.
Before you can read the data in memory, you must stop the memory
operations using the
SENSe:DIGital:MEMory:ENABle OFF
command.
Read all the captured data using the
SENSe:DIGital:MEMory:DATA:ALL?
query. This performs a non- destructive read of all data in the bank’s
memory.
To read specific captures, use the
SENSe:DIGital:MEMory:DATA?
form of
the command. This command takes
index
and
count
parameters to specify
which data to retrieve. The oldest data in memory has an
index
of ‘0’.
The
count
specifies the number of samples to read.
count
+
index
must
be less than the number of captured points.
Both these data reads are non- destructive to the bank memory.
To clear the memory for new data, send the
SENSe:DIGital:MEMory:CLEar
command.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...