Features and Functions
2
34980A User’s Guide
47
Scan Trigger Source
You can configure the event or action that controls the onset of each
sweep through the scan list (a
sweep
is one pass through the scan list):
•
You can set the instrument’s internal timer to automatically scan at a
specific interval. You can also program a time delay between channels
in the scan list (see
“Channel Delay”
on page 56).
•
You can manually control a scan by repeatedly pressing the
Scan (Measure)
key from the front panel.
•
You can start a scan by sending a software command from the remote
interface (
MEASure?
or
INITiate
command).
•
You can start a scan when an external TTL trigger pulse is received.
•
You can start a scan when an alarm event is logged on the channel
being monitored.
Interval Scanning
In this configuration, you control the frequency of scan sweeps by
selecting a wait period from the start of one trigger to the start of the
next trigger (called the
trigger- to- trigger interval
). If the scan interval
is less than the time required to measure all channels in the scan list, the
instrument will scan continuously, as fast as possible (no error is generated).
•
You can set the scan interval to any value between 0 seconds and
99:59:59 hours (359,999 seconds), with 1 ms resolution.
•
Once you have initiated the scan, the instrument will continue scanning
until you stop it or until the trigger count is reached. See
“Trigger
Count”
on page 52 for more information.
•
Mx+B scaling and alarm limits are applied to measurements during a
scan and all data is stored in volatile memory.
Trigger-to-trigger interval
t
Sweep 1
Trigger 1
Sweep 2
Sweep
n
Trigger 2
. . .
Trigger Timer
(0 to 359,999 seconds)
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...