Features and Functions
2
34980A User’s Guide
43
Scanning
The instrument allows you to combine a DMM (either internal or external)
with multiplexer channels to create a
scan
. During a scan, the instrument
connects the DMM to the configured multiplexer channels one at a time
and makes a measurement on each channel.
Any channel that can be “read” by the instrument can also be included in
a scan. This includes any combination of temperature, voltage, resistance,
current, frequency, or period measurements on multiplexer channels.
A scan can also include a read of a digital channel or a read of the
totalizer count on the digital modules. Scanning is allowed with the
following modules:
•
34921A through 34925A Multiplexer Modules
•
34950A Digital I/O Module (digital input and counter channels only)
•
34952A Multifunction Module (digital input and totalizer channels only)
Automated scanning is not allowed with the other switching modules.
In addition, a scan cannot include a write to a digital channel or a voltage
output from a DAC channel. You can, however, write your own program to
manually create a “scan” to include these operations.
Rules for Scanning
•
Before you can initiate a scan, you must set up a scan list to include all
desired multiplexer or digital channels. Channels which are not in the
scan list are skipped during the scan. By default, the instrument scans
the list of channels in ascending order from slot 1 through slot 8
(channels are reordered as needed). If your application requires
non- ordered scanning of the channels in the present scan list, see
“Non- Sequential Scanning”
on page 60. Measurements are taken only
during a scan and only on those channels which are included in the
scan list.
•
You can store at least 500,000 readings in memory and all readings are
automatically time stamped. If memory overflows, a status register bit
is set and new readings will overwrite the first (oldest) readings stored.
The most recent readings are always preserved. You can read the
contents of memory at any time, even during a scan. Reading memory
is not cleared when you read it.
•
Each time you start a new scan, the instrument clears all readings
(including alarm data) stored in reading memory from the previous
scan. Therefore, the contents of memory are always from the most
recent scan.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...