92
34980A User’s Guide
2
Features and Functions
•
You can display a message on the front panel by sending a command
from the remote interface. The instrument can display up to 18
characters on the upper line of the front- panel display; any additional
characters are truncated (no error is generated). You can use letters
(A- Z), numbers (0- 9), and special characters like “@”, “%”, “*”, etc.
Use the “#” character to display a degree symbol ( ° ). Commas, periods,
and semicolons share a display space with the preceding character, and
are not considered individual characters.
•
While a message is displayed on the front panel, readings from a scan
or monitor are not sent to the front- panel display.
•
Sending a text message to the display overrides the display state;
this means that you can display a message even if the display is
turned off. In addition, pressing any front- panel key will clear the
text message.
Remote Interface Operation:
The following command turns off the
front- panel display.
DISPLAY OFF
The following command displays a message on the front panel and turns
on the display if currently disabled (the quotes are not displayed).
DISPLAY:TEXT "SCANNING ..."
To clear the message displayed on the front panel (without changing the
display state), send the following command.
DISPLAY:TEXT:CLEAR
Front-Panel Number Format
The instrument can show numbers on the front- panel display with periods
or commas for the decimal point (radix) and thousands separator.
This feature is available from the front panel only.
•
The number format is stored in non- volatile memory, and does
not
change when power has been off, after a Factory Reset (
*RST
command), or after an Instrument Preset (
SYSTem:PRESet
command).
•
When shipped from the factory, a period is used as the radix character
and commas are used for the digits separator (e.g., +1.234,56 VDC).
Front Panel Operation:
Utility > MISC. SETTINGS > RADIX|THOUSAND SEPARATOR
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...