64-Bit Digital I/O Module with Memory and Counter
10
34980A User’s Guide
271
The following handshake command sets the synchronous handshaking
mode for the channels in bank 1.
CONF:DIG:HAND SYNC, (@1101)
This form of the handshaking command also allows you to optionally set
the input threshold, output drive level, and polarity of all the handshake
lines. For example, the following command sets bank 2 to use synchronous
handshaking, with an input threshold of 2.5 V, an output drive level of
2.5 V, and normal polarity. Other parameters such as the handshake
timing are set to default values (refer to the
Programmer’s Reference
Help file
for details
)
.
CONF:DIG:HAND SYNC, 2.5, 2.5, NORM, (@1201)
You can set parameters by using a sequence of commands instead of the
CONFigure
macro command. For example, the following command sequence
sets the handshaking mode to synchronous, the output drive to open
collector, and the handshake rate to 1 MHz.
CONF:DIG:HAND:MODE SYNC, (@1101)
CONF:DIG:HAND:DRIV OCOL, (@1101)
CONF:DIG:HAND:RATE 1000000, (@1101)
Setting the Handshake Line Parameters
You can set the handshake lines’ input threshold, output drive mode, and
output drive voltage. These settings affect all the handshake lines in the
bank. Handshake line polarity can be set for each individual handshake
line.
For example, you can invert the polarity of the handshake line H1 on
bank 2 with the following command.
CONF:DIG:HAND:POL INV, H1, (@1201)
You can set the output drive mode, output voltage, and input threshold for
all handshake lines in each bank. For example, the following commands set
the drive mode to active, the drive voltage to 4.5 V, and the input
threshold to 1.0 V on bank 2.
CONF:DIG:HAND:DRIV ACT, (@1201)
SOUR:DIG:HAND:LEV 4.5, (@1201)
SENS:DIG:HAND:THR 1, (@1202)
N O T E
The settings for drive mode, output drive level, and input threshold also
apply to the bank’s interrupt line.
N O T E
When using external pull-ups in the open collector mode, the outputs will
not exceed 5 V.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...