86
34980A User’s Guide
2
Features and Functions
Reading the List of Stored Sequences
From the remote interface only, you can read the names of all sequences
currently stored in memory.
•
When stored in memory, the user- defined sequence names are
converted to all uppercase letters. For example, when stored “MySeq_1”
is converted to “MYSEQ_1”.
•
Up to 500 unique sequences can be stored in non- volatile memory.
Each sequence is limited to 1024 bytes.
Remote Interface Operation:
The following command returns a
comma- separated list of sequence names currently stored.
ROUT:SEQ:CAT?
The above command returns a string in the form:
MYSEQ_1,PATH_DUT1,SW_PATH2
If no sequence names have been stored, a null string (“ ”) string
is returned.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...