34980A User’s Guide
2
Features and Functions
Clearing 34980A Memory
For security reasons, you may want to clear memory in the 34980A.
To clear all measurement results from memory, either cycle power to the
34980A or send the
*RST
command. This will also clear the internal DMM
settings and all channel configurations, Mx+B scaling constants, and all
alarm settings.
The following settings are stored in
non- volatile
memory:
•
Optional channel labels
•
Real- time system clock setting
•
Front- panel number format setting
•
GPIB address setting
•
LAN settings
•
Stored instrument states
To clear the stored instrument states, use the
MEMory:STATe:DELete:ALL
command. To clear non- volatile memory, with the exception of the
LAN MAC address and USB ID, use the
SYSTem:SECurity:IMMediate
command.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...