Self-Test
A
power- on
self- test occurs automatically when you turn on the
instrument. This limited test assures you that the instrument and all
installed plug- in modules are operational. This self- test does
not
perform
the extensive self test described below.
A
complete
self- test actually performs a series of internal tests and takes
approximately 20 seconds to execute. If all tests pass, you can have high
confidence that the instrument and all installed plug- in modules are
operational.
This feature is available from the remote interface only.
•
If you have a 34951A Isolated DAC Module installed, the complete
self- test will require an additional 15 seconds to complete per DAC
module (a memory test is performed).
•
The complete self- test will abort if any signals are connected to ABus1
via the rear- panel Analog Bus connector (pins 4, 5, and 9; see
“Analog
Buses”
on page 16). Be sure to disconnect any signals from ABus1 prior
to running the self- test.
•
If the power- on or complete self- test fails, and error is stored in the
error queue. See the Agilent 34980A
Service Guide
for more
information on returning the instrument to Agilent for service.
•
Following the complete self- test, the instrument issues a Factory Reset
(
*RST
command).
Remote Interface Operation:
The following command returns “+0” if the
self- test is successful or “+1” if it fails.
*TST?
Front-Panel Display Control
For security reasons or for a slight increase in measurement rates,
you may want to turn off the front- panel display. From the remote
interface, you can also display up to 18 characters on the upper line
of the front- panel display.
•
You can disable the front- panel display only by sending a command
from the remote interface (i.e., you cannot disable the front panel while
in local operation).
•
When disabled, the entire front- panel display goes dark and all display
annunciators except ERROR, HOT, and Safety Interlock are disabled.
•
The front- panel display is automatically enabled when power is cycled,
after a Factory Reset (
*RST
command), or after an Instrument Preset
(
SYSTem:PRESet
command).
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...