General Purpose Switch Modules
6
34980A User’s Guide
189
After a five- to ten- second delay, remove the sheet metal cover from the
module and move the position of the jumper mounted on the module.
See the figure below for the jumper’s location on the module.
WA RN I N G
Before changing the position of the jumper, remove external
connections from the module. Wait five to ten seconds to allow
the module’s internal capacitors to discharge.
WA RN I N G
Do not connect either the 34937A or 34938A module directly to a
mains power outlet. If it is necessary to switch a mains voltage or
any circuit where a large inductive load may be switched, you
must add signal conditioning elements to reduce the potential
transients before they reach the module or the Analog Buses.
U205
U301
C301
Open
Maintain
5Amp relays
Power Down State
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...