Low Frequency Multiplexer Switch Modules
4
34980A User’s Guide
153
Four-Wire
•
one 20- channel 4- wire MUX. This configuration requires using neither
external wiring nor connecting through the internal Analog Buses. For
4- wire resistance measurements, the instrument automatically pairs
channel
n
on Bank 1 with channel
n
+20 (Bank 2) to provide the
source and sense connections. Four- wire controls occur only when
doing 4- wire measurement operations through the internal DMM, such
as
MEASure:FRESistance?
or scanning a channel previously
configured as 4- wire.
One-Wire
•
two independent 40- channel 1- wire MUXes. This configuration
requires neither using external wiring nor connecting through the
Analog Bus relays.
•
one 80- channel 1- wire MUX. You must use external wiring or connect
through the Analog Bus relays for this configuration.
Interlock Protection
This module is interlock protected, which means whenever the D- sub
connector end of the modules is exposed, the Analog Bus relays
immediately open and disconnect from the Analog Buses. For more
information, refer to
page 120
.
Overvoltage Protection
This module also features high voltage detection (< 100 V) and current
limiting circuitry to protect the FET relays. This circuitry senses current
flows from input overvoltages. These overvoltages may come from either
the MUX input or from the Analog Buses. In addition, each channel is
also protected from input overvoltages with a resistor.
When overvoltage is detected, all relays (Analog Bus and FET) are
opened. While in the overvoltage state, any attempts to close any Analog
Bus or FET switch, results in an error status response from the module.
Once in the overvoltage state, you must restore normal module operation
with one of these actions:
•
using the
SYSTem:CPON <slot>
command. This affects only the
module specified.
N O T E
Because all bank relays supply only HI signals, you can apply a
LOW signal through COM1 L or COM2 L when you are making
2-wire resistance measurements in 1-wire mode.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...