RF Multiplexer Switch Modules
7
34980A User’s Guide
203
Example: Clearing the cycle count for a relay
The following command resets
the cycle count on the channels 103 and 201 for a module in slot 1.
Note that clearing the cycle count on a specific channel will clear the
count on
all
three relays in the corresponding bank.
DIAGnostic:RELay:CYCLes:CLEar (@1103,1201)
Example: Resetting module to power-on state
The following command resets
a module in slot 4 to its power- on state.
SYSTem:CPON 4
34941A and 34942A Simplified Schematic
Both the 34941A and 34942A modules are configured alike. They each
contain four banks of latching switches. Each bank consists of three form
C relays.
The front panel of the two RF MUX modules are similar with channel
labels in the same positions, the unique product number on the left, and
the product description on the right.
301
302
303
304
404
403
402
401
COM
COM
101
102
103
104
204
203
202
201
COM
COM
3494xA
Quad 1x4
RF MUX
xx Ohm
Bank 1
Bank 2
Bank 3
Bank 4
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...