Features and Functions
2
34980A User’s Guide
49
Remote Interface Operation:
The following program segment configures the
instrument for a manual scanning operation.
Then, send the
*TRG
(trigger) command to begin each scan sweep.
The
*TRG
command will not be accepted unless the internal DMM is in
the “wait- for- trigger” state.
Note:
To stop a scan, send the
ABORt
command.
Scanning on Alarm
In this configuration, the instrument initiates a scan each time a reading
crosses an alarm limit on a channel. You can also assign alarms to
channels on the digital modules (34950A and 34952A). For example,
you can generate an alarm when a specific bit pattern or bit pattern
change is detected on a digital input channel or when a specific count is
reached on a totalizer channel.
•
In this scan configuration, you may use the Monitor function to
continuously take readings on a selected channel and wait for an alarm
on that channel. Channels do not have to be part of an active scan list
to be monitored; however, the channel must be configured for a
measurement in order to be monitored.
•
All readings from the scan are stored in volatile memory.
Readings accumulate in memory until the scan is terminated
(until the trigger count is reached or until you abort the scan).
•
You can specify a trigger count which sets the number of front- panel
key presses or scan trigger commands that will be accepted before
terminating the scan. See
“Trigger Count”
on page 52 for more
information.
•
Mx+B scaling and alarm limits are applied to measurements during a
manual scanning operation and all data is stored in volatile memory.
TRIG:SOURCE BUS
TRIG:COUNT 2
INIT
Select bus (manual) mode
Sweep the scan list 2 times
Initiate the scan
N O T E
For complete details on configuring and using alarms, refer to
“Alarm Limits”
on page 68.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...