202
34980A User’s Guide
7
RF Multiplexer Switch Modules
34941A and 34942A SCPI Programming Examples
The programming examples below provide you with SCPI command
examples to use for actions specific to the RF MUX switch modules.
The slot and channel addressing scheme used in these examples follow
the form
sccc
where
s
is the mainframe slot number (1 through 8) and
ccc
is the channel number. For information on specific configurations,
refer to the simplified schematic on
page 203
.
For complete information on the SCPI commands used to program the
34980A, refer to the Agilent 34980A Programmer’s Reference contained
on the 34980A Product Reference CD. For example programs, also refer
to the 34980A Product Reference CD.
Example: Closing channels
You can only close channels on the RF MUX
modules. You cannot open channels. When you close a channel, any
already- closed channels automatically open. With this “one- step”
operation, the relays switch in the proper order that avoids momentary
connection of the wrong input to the multiplexer output. The following
command closes channel 03 on Bank 1 of an RF MUX module installed
in slot 5.
ROUTe:CLOSe (@5103)
Example: Querying channels for open or close state
The following commands
returns the close or open state of channel 33 of a module installed in
slot 5.
ROUT:CLOSe? (@5033)
ROUT:OPEN? (@5033)
Example: Querying the system for module identify
The following command
returns the identify of the module installed in slot 7.
SYSTem:CTYPe? 7
Example: Reading the cycle count for a relay
On these modules, each bank
consists of two
leaf
relays and one
tree
relay (see the simplified
schematic on
page 203
). The module stores the cycle count for each of
the three relays on all four banks. The cycle count is the greater of the
three values on the specified bank (i.e., reflecting the cycle count for the
entire bank). Therefore, the count for Channels 101, 102, 103, and 104
will
always
be equal. The following statement reads back the number of
completed cycles for the channels 101 and 202 on a module installed in
slot 6.
DIAGnostic:RELay:CYCLes? (@6101,6202)
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...