6-12
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Screen Organization
CHANNELS Screen Group
CHANNELS: USER FUNCTION DEFINITION screen
On this screen, you define user functions. For details about user functions, refer to
“User Function” in Chapter 7.
User Comment
In this field, you can enter a desired comment, which is also displayed in the
CHANNELS, MEASURE, DISPLAY, and GRAPH/LIST screen groups.
NAME
NAME field defines the user function name. In this field, you can enter a name by
using the keyboard or front panel keys. Or you can select variables that are shown
on the secondary softkeys.
To delete a user function, you can select DELETE ROW softkey to delete the
NAME, UNIT, and DEFINITION entries.
After defining a user function, you can use this variable name for reference on other
screens.
Restrictions
•
NAME must be 6 or less alphanumeric characters. First character must be
alphabet character.
•
NAME must be different from other names. The alphabet characters are case
sensitive. For example, HFE is different from Hfe.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...