Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
6-27
Screen Organization
MEASURE Screen Group
•
DELAY TIME
DELAY TIME field specifies the delay time from the pulse period start time.
The delay time must be less than or equal to the pulse period.
•
PEAK VALUE and BASE VALUE
PEAK VALUE and BASE VALUE fields specify the pulse peak and pulse base
values.
•
LEADING TIME and TRAILING TIME
LEADING TIME and TRAILING TIME fields specify the transition time of
leading and trailing edges, which is time for pulse to change from 10% to 90%
of pulse amplitude.
•
IMPEDANCE
IMPEDANCE field specifies the PGU output impedance. In this field, select:
•
LOW secondary softkey to set output impedance to about 0
W
.
•
50 ohm secondary softkey to set output impedance to 50
W
.