Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
2-9
Measurement Mode
Sweep Measurement Mode
Synchronous Sweep Measurement
For synchronous sweep measurement, you set up a synchronous sweep source
(VAR1') in addition to a primary sweep source (VAR1). The output of the
synchronous sweep source is swept synchronously with the output of the primary
sweep source at a constant offset value and ratio.
Figure 2-3
Synchronous Sweep Measurement
To set up synchronous sweep measurement, select the following on CHANNELS:
CHANNEL DEFINITION page:
•
VAR1 function for desired primary sweep source (SMU or VSU).
•
VAR1' function for desired synchronous sweep source (SMU or VSU).
Synchronous sweep mode has the following restrictions:
•
For the following, VAR1' is always set to the same mode as VAR1:
•
linear/log staircase
•
single/double sweep mode
•
VAR1 and VAR1' must be same V/I output mode. For example, if VAR1 is set to
V
mode, then VAR1' must be set to
V
or
VPULSE
mode.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...