3-32
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Operation States
Stress Force State
In the stress force state, the 4155B/4156B outputs stress. The output switches are off
for units that do
not
have entries in the CHANNELS table of the STRESS:
CHANNEL DEFINITION screen.
Standby State
In the standby state, the 4155B/4156B does not perform measurements or stress
force, but it outputs dc bias and/or pulses using the measurement units defined as the
standby channel. Following settings keep the same conditions as the previous state:
•
output switch of each unit
•
output impedance of PGUs
•
switching condition of the 16440A selector
•
resistance selection of the 16441A R-Box
Changing among Operation States
Figure 3-3 shows how to change among the operation states.
•
From measurement/stress states to idle state:
If you perform measurements or force stress from the idle state, then the
4155B/4156B returns to the idle state after one of the following conditions
occurs:
•
Measurement is finished.
•
Fixture lid is opened while an SMU is outputting more than ±40 V.
•
Stress is finished.
•
Stop front-panel key is pressed.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...