Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-51
Measurement Functions
R-BOX Control
Circuit Diagram
Figure 3-8 shows a simplified circuit diagram of an 16441A R-Box.
Table 3-4 shows switching conditions for each setting.
Table 3-4
Switching Conditions of the 16441A R-Box
Resistance is switched before and after measurement state. In the standby state, the
stress state, and the idle state, 0
W
is connected.
Settings
Switches
SW1 SW2 SW3 SW4
0
W
ON
ON
OFF
OFF
10 k
W
OFF
OFF
OFF
ON
100 k
W
OFF
OFF
ON
OFF
1 M
W
OFF
OFF
OFF
OFF