6-76
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Screen Organization
STRESS Screen Group
V
dc voltage source
I
dc current source
VPULSE
ac voltage source
COMMON
circuit common
DELETE ROW
Deletes all entries in row of unit, so unit is not used during
stress force. Output switch of unit is open.
Allowable modes for each unit are shown in the following table:
When pointer is at
top of this column
, CHANNEL ASSIGN softkey appears:
NOTE
Switching units
To switch the STRESS MODE, NAME, and FCTN assignments for two units,
do as follows:
1. Position pointer in top field of STRESS MODE column. CHANNEL
ASSIGN softkey appears.
2. Select CHANNEL ASSIGN. Pointer moves to the top field of UNIT
column.
3. Use arrow keys in the MARKER/CURSOR key group to move pointer to
desired row.
4. Select the secondary softkey of the desired unit. The selected unit appears at
the pointer.
Perform steps 3 and 4 until you assign units as desired. Make sure that the same unit
is not assigned to multiple rows. Then select EXIT CHANNEL ASSIGN softkey.
V I
VPULSE
COMMON
DELETE
ROW
SMU
yes
yes
yes
yes
VSU
yes
yes
PGU
yes
yes
yes
GNDU
yes
yes
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...