Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
4-21
Making a Measurement
Sweep Measurements
To Set up SMU Pulsed Output
SMU pulse output source is the measurement unit defined as VPULSE or IPULSE
in the CHANNELS: CHANNEL DEFINITION screen. For pulsed sweep source, set
the function (FCTN) to VAR1, VAR2, or VAR1’. For pulsed constant source, set the
function to CONST. To set up the SMU pulse output source, press Meas front-panel
key. The MEASURE: SWEEP SETUP screen is displayed.
1. Set the source output parameters:
•
For VAR1, refer to “To Set up Primary Sweep” on page 4-14.
•
For VAR2, refer to “To Set up Secondary Sweep” on page 4-16.
•
For VAR1’, refer to “To Set up Synchronous Sweep” on page 4-18.
•
For CONST, refer to “To Set up Constant Output” on page 4-20.
2. SMU PULSE: PERIOD
Enter the pulse period value.
3. SMU PULSE: WIDTH
Enter the pulse width value.
4. SMU PULSE: BASE
Enter the pulse base value.
You cannot change UNIT and NAME in this screen. To change the values, go to the
CHANNELS: CHANNEL DEFINITION screen.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...