
3-40
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Trigger Function
Setup and restrictions
•
You
cannot
perform trigger outputs together with trigger inputs. You
must
select
either trigger output or trigger input.
•
To use a trigger function, you must enable the trigger function and select either
TRIG OUT or TRIG IN in the TRIGGER SETUP table on the MEASURE:
OUTPUT SEQUENCE screen. Then the trigger inputs or outputs are performed
automatically after you start a measurement by selecting a measurement
front-panel key (Single, Repeat, or Append).
•
The
trigger output
function is
not
available for sampling measurements.
•
When you perform knob sweep measurements, the trigger function is not
available.
•
For the electrical specifications of trigger signals, refer to
User's Guide General
Information
.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...