Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
6-19
Screen Organization
MEASURE Screen Group
VAR2 parameters
In this column, you set up the output parameters for the secondary sweep unit. UNIT
and NAME are defined on the CHANNELS: CHANNEL DEFINITION screen.
SWEEP MODE and LIN/LOG fields are fixed to SINGLE and LINEAR.
•
START, STEP, and NO. OF STEP
In the START, STEP, and NO OF STEP fields, you specify the start value, step
value, and number of steps. The stop value is calculated from these values, and
is shown in the STOP field.
•
COMPLIANCE
In COMPLIANCE field, you set compliance value. If a VSU is used for VAR2
unit, this field
cannot
be set: compliance value is fixed to approximately 100
mA.
•
POWER COMP
In POWER COMP field, you can set power compliance value for SMUs. To
disable power compliance function, select OFF secondary softkey.
VAR1' parameters
In this column, you set up the output parameters for the synchronous sweep unit.
This VAR1' table is displayed only when VAR1' is set in the FCTN field on the
CHANNELS: CHANNEL DEFINITION screen.
UNIT and NAME are defined on CHANNELS: CHANNEL DEFINITION screen.
•
OFFSET and RATIO
In the OFFSET and RATIO fields, you specify the
offset
and
ratio
values. The
offset and ratio values determine the VAR1' value as follows:
VAR1’ output = VAR1 output
´
ratio
+
offset
•
COMPLIANCE
In COMPLIANCE field, you set compliance value. If a VSU is used for VAR1'
unit, this field
cannot
be set: compliance value is fixed to 100 mA.
•
POWER COMP
In the POWER COMP field, you can set the power compliance value. To disable
the power compliance function, select OFF secondary softkey. If
an
SMU is set
to VPULSE or IPULSE mode and if the SMU is set to VAR1', you
cannot
set
power compliance for the VAR1' SMU.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...