2-6
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Mode
Sweep Measurement Mode
start
Start value of sweep. For logarithmic sweep,
start
must not be
zero. Allowable range of
start
depends on output range of
sweep source. For output range of each measurement channel,
refer to Chapter 1.
stop
Stop value of single sweep or turning back value of double
sweep. For logarithmic sweep,
stop
must have same polarity as
start
, and must not be zero. Allowable range of
stop
depends on
output range of sweep source. For output range of each
measurement channel, refer to Chapter 1.
step
•
For linear sweep,
step
is step increment of sweep. Number
of sweep steps is calculated from
start
,
stop
, and
step
.
Calculated number of steps must be in range: 2 to 1001.
•
For logarithmic sweep,
step
is invalid. Number of sweep
steps is calculated from
start
,
stop
, and number of steps in
one decade, which is specified by
log
parameter. Calculated
number of steps must be in range: 2 to 1001.
compliance
Compliance value of sweep source. This parameter applies to
SMU only. Allowable range of
compliance
depends on the
compliance range of sweep source. For the compliance range of
each measurement channel, refer to Chapter 1.
power compliance
(Optional) Power compliance value of sweep source. This
parameter applies to SMU only. Allowable range depends on
power compliance range of sweep source. For details, refer to
Chapter 3.
hold time
Time required for DUT to settle after forcing start value.
Allowable range is 0 to 655.35s. Resolution: 10 ms.
delay time
Time required for DUT to settle after stepping the output.
Allowable range: 0 to 65.535 s. Resolution: 100
m
s
Refer to “CHANNELS: CHANNEL DEFINITION screen” and “MEASURE:
MEASURE SETUP screen” in Chapter 6 for setting up these parameters.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...