Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
2-3
Measurement Mode
Sweep Measurement Mode
Sweep Measurement Mode
For sweep measurements, the sweep source channels perform staircase sweep
output of voltage or current, while the monitor channels measure voltage or current
for each sweep step.
Only SMUs and VSUs can be sweep sources (VAR1, VAR2, and VAR1').
The 4155B/4156B provides three types of sweep measurement:
•
One sweep source (VAR1) is used.
•
“Subordinate Sweep Measurement”
A primary (VAR1) and secondary sweep source (VAR2) are used.
•
“Synchronous Sweep Measurement”
A primary (VAR1) and synchronous sweep source (VAR1') are used.
Also, you can set up a combined subordinate and synchronous sweep measurement.
In addition to the normal dc sweep, the sweep or constant source output can be
pulsed to prevent thermal drift of the DUT.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...