Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-65
Measurement Functions
SMU Filter
SMU Filter
You can set SMU filter to on or off for sampling measurements or stress forcing. If
filter is
on
, noise and overshoot are decreased, but settling time takes longer.
•
sampling measurement
You set the FILTER field on the MEASURE: SAMPLING SETUP screen.
If you set initial interval to a short time, and if filter is set to
ON
, be aware that
settling time takes several ms.
•
stress force
You set the FILTER field on the STRESS: STRESS SETUP screen.
If you set dc stress to short stress force time, set
OFF
in this field if you want the
stress signal to be more pulsed shaped.
•
sweep measurement
When you perform sweep measurements, the SMU filter conditions are
automatically set as follows:
For a pulsed SMU
Filter is
off
.
For non-pulsed SMUs
Filters are
on
.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...