Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
5-29
Analyzing Measurement Results
Automatic Analysis
Data condition mode specifies a point related to the measurement curve. So, if no
measurement data satisfy the specified condition, the nearest measurement point is
used.
For the meaning of expression that you can enter in step 6 or 7, see “Expression” in
Chapter 7.
To specify a point between two measurement points
Set Interpolate field to
ON
.
To disable (clear) the settings
Move the pointer to field (1), then select DISABLE secondary softkey. Setup fields
disappear.
Example
The following figure shows an example setup to automatically draw a line through
two specified points. One point is specified by X-Y coordinate mode and other point
is specified by data condition mode.