8-6
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
If You Have A Problem
When You Make A Measurement
If Measured Value Oscillates when Measuring Negative
Resistance
If the DUT has negative resistance characteristics, SMUs may oscillate. Because
SMUs operate as negative feedback amplifier.
To solve this problem:
•
For voltage controlled negative resistance device
•
Connect G in parallel with your DUT to cancel negative resistance. To
obtain an output I-V curve, use the following equation.
I
Y
= I
-
G * V
•
For current controlled negative resistance device
•
Connect R in series with your DUT to cancel negative resistance. To obtain
an output I-V curve, use the following equation.
V
Z
= V
-
R * I
•
If the resistance of the DUT is less than 1 M
W
, you can use R-Box.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...