3-64
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Integration Time
Medium
Medium integration time is 1 PLC (power line cycle).
Setting
Press Medium front-panel key.
Integration time
depends on the power line cycle. cannot change this value.
(for example, 20 ms for 50 Hz)
If you measure current in the 1 nA or lower ranges by using SMUs, integration time
of SMUs is automatically changed as follows:
Long
Long integration time is effective when you need high resolution and noise
reduction measurement. But the measurement speed is slow.
Setting
Press Long front-panel key.
Integration time
2 PLC to 100 PLC with 1 PLC resolution
Initial setting
16 PLC
When an 4155B/4156B measures current in 1 nA or lower ranges by using HRSMU,
integration time of HRSMU is automatically changed to longer integration time
(maximum 100 PLC) than specified.
Measurement Unit
Measurement Range
Integration Time
HRSMU
10 pA
50 PLC
100 pA
10 PLC
1 nA
5 PLC
MPSMU
1 nA
3 PLC
HPSMU
1 nA
3 PLC
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...