4-38
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Knob Sweep Measurements
Warning messages
If the CHANNELS or MEASURE screen group have incorrect settings for knob
sweep measurements when starting the knob sweep, a warning message is
displayed, then the STOP and CONT primary softkeys are available.
Select STOP softkey to know the incorrect setting. The setting is highlighted on the
screen.
Select CONT softkey to perform knob sweep measurement with the following
settings:
•
If VAR1’ function (FCTN) has been set:
The unit works as CONST channel. The output value is VAR1’ start value.
•
If VPULSE or IPULSE mode (MODE) has been set:
The unit works as V or I channel.
•
If the power compliance function (POWER COMP) has been set:
The function does not work. Same as POWER COMP = OFF.
Example
The following figure shows an example to set both X axis and Y axis display
regions to positive.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...