If You Have A Problem
When You Make A Measurement
If Noise Affects the Measured Values
When you measure low current of a DUT, the measured values may not be stable.
To solve this problem:
•
Use guarding to reduce the leakage current between your prober and the
4155B/4156B. Note that long wires cause oscillation because of their large
inductance. For details about connections, refer to “To Make Connections to
Reduce Leakage Current” in Chapter 4.
•
If some high-power electric machines are operating around the 4155B/4156B,
turn off the machines, then perform the measurements. The machines affect the
power line waveform.
•
Shut the lid of test fixture or shield box to prevent effects of light.
•
If these are vibrations due to nearby machines or due to air flow, put cushioning
material under prober, cable, and the 4155B/4156B; install stabilizer on the
prober; and make the cables stable by taping.
•
Wait several minutes after connecting cables or moving probe needles. Because
these operations cause electromotive force.
•
If you use only Force terminal and triaxial cables for HRSMUs or HPSMU,
connect an open cap to sense terminal.
•
Keep constant temperature in the room when you use the 4155B/4156B. Shift of
1
°
C may shift the measurement values. Temperature change causes the
following.
•
Offset current in the 4155B/4156B.
•
Thermoelectromotive force in DUT, which causes low current.
•
Expansion and contraction of cables, which causes noise.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...