Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
4-47
Making a Measurement
Sampling Measurements
To Define Stop Conditions
The measurement stop condition defines the condition to stop the sampling
measurement. The stop condition is one of the sampling completion conditions. For
the sampling completion conditions, see “Sampling Measurement Mode” in Chapter
2.
1. Press Meas key in the PAGE CONTROL key group.
2. Select SAMPLNG SETUP primary softkey.
3. In the ENABLE/DISABLE field of the STOP CONDITION area, select
ENABLE secondary softkey.
4. In NAME field of STOP CONDITION area, select the secondary softkey for the
desired variable name or user function name to be used for EVENT comparison.
5. In THRESHOLD field of STOP CONDITION area, enter the threshold value for
the name selected in the previous step.
6. In EVENT field of STOP CONDITION area, select:
•
Val > Th secondary softkey to stop the sampling when the sampled value is
greater than the threshold value.
•
Val < Th secondary softkey to stop the sampling when the sampled value is
less than the threshold value.
•
|Val| > |Th| secondary softkey to stop the sampling when the absolute
sampled value is greater than the absolute threshold value.
•
|Val| < |Th| secondary softkey to stop the sampling when the absolute
sampled value is less than the absolute threshold value.
7. In EVENT NO. field, enter a value which specifies the sampling to stop when
EVENT occurs EVENT NO. times.
If you select DISABLE in the ENABLE/DISABLE field, the sampling
measurement continues until:
•
Stop key in the MEASUREMENT key group is pressed.
•
Specified total sample time has elapsed.
•
The 4155B/4156B receives GPIB command to stop sampling.
•
An emergency condition occurs on the 4155B/4156B.
•
Interlock terminal opens due to high voltage. (See Chapter 2 of
User's Guide
General Information
.)
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...