4-46
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Sampling Measurements
To Set up Constant Output
1. Define CONST units as described in “To Define Measurement Units” on page
2. Press Meas key in the PAGE CONTROL key group.
3. Select SAMPLNG SETUP primary softkey.
4. In the SOURCE field of the desired unit in the CONSTANT area, enter the
desired output value.
To modify the UNIT, NAME, and MODE field
Modify the UNIT, NAME, and MODE fields on the CHANNELS: CHANNEL
DEFINITION screen.
To set up compliance value for constant output
Set desired value in the COMPLIANCE field of the CONSTANT table. For details
about compliance, see “Compliance” in Chapter 3.
Example
The following example shows the constant output conditions: