4-18
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Sweep Measurements
To Set up Synchronous Sweep
Synchronous sweep source is the measurement unit defined as VAR1’ in the
CHANNELS: CHANNEL DEFINITION screen. VAR1’ is available for the
measurement units set to the output mode same as the VAR1 output mode. The
output value of VAR1' is calculated by the following equation:
VAR1' = VAR1
´
RATIO + OFFSET
On the MEASURE: SWEEP SETUP screen, set up the primary sweep source
(VAR1), then do following:
1. VAR1’ : OFFSET
Enter the offset value of the VAR1’ output against the VAR1 output.
2. VAR1’ : RATIO
Enter the ratio value of the VAR1’ output against the VAR1 output.
3. VAR1’ : COMPLIANCE, POWER COMPLIANCE
Only for SMU. Enter the compliance value, and power compliance (ON or OFF)
for the synchronous sweep source.
You cannot change UNIT and NAME in this screen. To change the values, go to the
CHANNELS: CHANNEL DEFINITION screen.
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...