2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Test Internal Trigger Sensitivity (all models)
1
On the oscilloscope, press the
[Default Setup]
key.
2
Press the
[Mode/Coupling]
key; then, press the
Mode
softkey to select
Normal
.
3
Connect the equipment (see
Figure 8
).
a
Connect the signal generator output to the oscilloscope
channel 1 input.
Figure 8
Setting Up Equipment for Internal Trigger Sensitivity Test
b
With 2000 X- Series oscilloscopes, connect a 50 ohm
feedthrough termination between the channel 1 input
and the BNC cable.
With 3000 X- Series oscilloscopes, set channel 1
Imped
to
50 Ohm
.
4
To verify the trigger sensitivity at the oscilloscope’s
maximum bandwidth, set the output frequency of the
Signal
Generator
Oscilloscope
N Cable
N to BNC Adapter