2000/3000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
1
Perform the procedure
“To perform an Auto Scale on the
Probe Comp signal”
on page 80
2
Press the channel key to which the probe is connected
(
[1]
,
[2]
, etc.).
3
In the Channel Menu, press
Probe
.
4
In the Channel Probe Menu, press
Probe Check
; then, follow
the instructions on- screen.
If necessary, use a nonmetallic tool (supplied with the
probe) to adjust the trimmer capacitor on the probe for
the flattest pulse possible.
On the N2862/63/90 probes, the trimmer capacitor is the
yellow adjustment on the probe tip. On other probes, the
trimmer capacitor is located on the probe BNC connector.
Figure 13
Example pulses
5
Connect probes to all other oscilloscope channels (channel
2 of a 2- channel oscilloscope, or channels 2, 3, and 4 of a
4- channel oscilloscope).
6
Repeat the procedure for each channel.
The process of compensating the probes serves as a basic
test to verify that the oscilloscope is functional.
comp.cdr
Perfectly compensated
Over compensated
Under compensated
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...