40
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
7
Adjust the output so that the multimeter reading displays
the first Volts/div calibrator setting value in
Table 8
or
Table 9
(depending on the oscilloscope model).
8
Disconnect the multimeter.
9
Wait until the measurement settles.
10
Read the “current” average voltage value again as V2.
11
Calculate the difference V2 - V1.
The difference in average voltage readings should be
within the test limits of
Table 8
or
Table 9
(depending on
the oscilloscope model).
If a result is not within the test limits, go to the
“Troubleshooting” chapter. Then return here.
12
Disconnect the oscilloscope calibrator from the
oscilloscope.
13
Repeat this procedure to check the DC vertical gain
accuracy with the remaining Volts/div setting values in
Table 8
or
Table 9
(depending on the oscilloscope model).
14
Finally, repeat this procedure for the remaining channels
to be tested.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...