![Agilent Technologies 2000 X Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_service-manual_2868574059.webp)
Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
59
signal generator to the maximum bandwidth of the
oscilloscope:
•
1 GHz models: 1 GHz.
•
500 MHz models: 500 MHz.
•
350 MHz models: 350 MHz.
•
200 MHz models: 200 MHz.
•
100 MHz models: 100 MHz.
•
70 MHz models: 70 MHz.
5
Perform these steps to test at the 5 mV/div setting:
a
Set the signal generator amplitude to about 10 mV
pp
.
b
Press the
[AutoScale]
key.
c
Set the time base to 10 ns/div.
d
Set channel 1 to 5 mV/div.
e
Decrease the amplitude from the signal generator until
1 vertical division of the signal (about 5 mV
pp
) is
displayed.
The trigger is stable when the displayed waveform is
stable. If the trigger is not stable, try adjusting the
trigger level. If adjusting the trigger level makes the
trigger stable, the test still passes. If adjusting the
trigger does not help, see the “Troubleshooting” chapter.
Then return here.
f
Record the result as Pass or Fail in the Performance
Test Record (see
page 63
).
g
Repeat this step for the remaining oscilloscope
channels.
6
Perform these steps to test at the 10 mV/div setting:
a
Set the signal generator amplitude to about 20 mV
pp
.
b
Press the
[AutoScale]
key.
c
Set the time base to 10 ns/div.
d
Set channel 1 to 10 mV/div.
e
Decrease the amplitude from the signal generator until
0.6 vertical divisions of the signal (about 6 mV
pp
) is
displayed.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...