Replacing Assemblies
5
2000/3000 X-Series Oscilloscopes Service Guide
117
To remove the acquisition board
These steps describes how to remove the acquisition board.
1
Using a TORK T6 driver, locate and remove the 4 screws
on the front of the instrument (4 Channel version).
Figure 37
Removing the T6 screws (4 channel version)