![Agilent Technologies 2000 X Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/agilent-technologies/2000-x/2000-x_service-manual_2868574076.webp)
76
2000/3000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
✔
Check that the trigger setup is correct.
✔
A correct trigger setup is the most important factor in
helping you capture the data you desire. See the
User’s
Guide
for information about triggering.
✔
Check that persistence in the Display menu is turned off,
then press the
Clear Display
softkey.
✔
Press the
[Auto Scale]
key.
If you cannot see a channel
✔
Recall the default setup by pressing
[Default Setup]
. This
will ensure that the trigger mode is Auto.
✔
Check that the oscilloscope probe’s BNC connector is
securely attached to the oscilloscope’s input connector.
✔
Check that the probe clips are securely connected to
points in the circuit under test.
✔
Check that the circuit under test is powered on.
You may have pressed the
[Auto Scale]
key before an input
signal was available.
Performing the checks listed here ensures that the signals
from the circuit under test will be seen by the
oscilloscope. Perform the remaining checks in this topic to
make sure the oscilloscope channels are on, and to obtain
an automatic setup.
✔
Check that the desired oscilloscope channels are turned
on.
a
Press the analog channel key until it is illuminated.
b
On models with the MSO option, press the digital
channels
[Digital]
key until it is illuminated.
✔
Press the
[Auto Scale]
key to automatically set up all
channels.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...