Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
47
9
Press the
Y2
softkey, then position the Y2 cursor to the
center of the voltage trace using the Entry knob.
The
Δ
Y value on the lower line of the display should be
within the test limits of
Table 11
or
Table 12
(depending
on the oscilloscope model).
If a result is not within the test limits, go to the
“Troubleshooting” chapter. Then return here.
10
Disconnect the oscilloscope calibrator from the
oscilloscope.
11
Repeat this procedure to check the dual cursor accuracy
with the remaining Volts/div setting values in
Table 11
or
Table 12
(depending on the oscilloscope model).
12
Finally, repeat this procedure for the remaining channels
to be tested.