2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
To verify digital channel threshold accuracy (MSO models only)
This test verifies the digital channel threshold accuracy
specification of the Agilent 2000/3000 X- Series oscilloscopes.
Threshold accuracy test limits: ±(100 mV + 3% of threshold
setting)
When to Test
You should perform this test every 12 months or after 2000
hours of operation, whichever comes first.
What to Test
Use these instructions to test the threshold settings of digital
channels D7- D0. Then, use the same instructions to test
digital channels D15- D8 (on 3000 X- Series).
Verifying Test Results
After each threshold test, record the voltage reading in the
Performance Test Record on
page 63
. To verify whether a
test passes, verify that the voltage reading is within the
limits in the Performance Test Record.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...