Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
43
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall
the factory default setup.
2
If you are testing a 2000 X- Series oscilloscope, set the
probe attenuation to 1:1 on the analog channel you are
testing (for example,
[1] > Probe > Probe
; then, turn the
Entry knob to select
1.00 : 1
).
3
Set up the oscilloscope.
a
Set the Volts/Div setting to the value in the first line in
Table 11
or
Table 12
(depending on the oscilloscope
model).
b
Adjust the channel 1 position knob to place the
baseline at 0.5 major division from the bottom of the
display.
Table 10
Equipment Required to Verify Dual Cursor Accuracy
Equipment
Critical Specifications
Recommended
Model/Part
Oscilloscope
Calibrator
14 mV to 35 Vdc,
0.1 V resolution
Fluke 5820A
Digital multimeter
Better than 0.01% accuracy
Agilent 34401A
Cable
BNC, Qty 2
Agilent 10503A
Shorting cap
BNC
Agilent 1250-0774
Adapter
BNC (f) to banana (m)
Agilent 1251-2277
Adapter
BNC tee (m) (f) (f)
Agilent 1250-0781 or
Pomona 3285
Blocking capacitor
Agilent
Pomona 4288 +
Pomona 5088
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...