2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Table 11
Settings Used to Verify Dual Cursor Accuracy, 2000 X-Series
Models
Volts/Div
Setting
Oscilloscope
Calibrator Setting
Test Limits
5 V/Div
35 V
33.6 V
to
36.4 V
2 V/Div
14 V
13.44 V
to
14.56 V
1 V/Div
7 V
6.72 V
to
7.28 V
500 mV/Div
3.5 V
3.36 V
to
3.64 V
200 mV/Div
1.4 V
1.344 V
to
1.456 V
100 mV/Div
700 mV
672 mV
to
728 mV
50 mV/Div
350 mV
336 mV
to
364 mV
20 mV/Div
140 mV
134.4 mV
to
145.6 mV
10 mV/Div
70 mV
67.2 mV
to
72.8 mV
5 mV/Div
1
35 mV
33.2 mV
to
36.8 mV
2 mV/Div
1, 2
14 mV
12.56 mV
to
15.44 mV
1 mV/Div
1, 2
7 mV
5.56 mV
to
8.44 mV
1
A blocking capacitor is required at this range to reduce noise. See
“Use a
Blocking Capacitor to Reduce Noise”
on page 48.
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...