Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
53
9
Change the signal generator output frequency according to
the maximum frequency for the oscilloscope using the
following:
•
1 GHz Models: 1 GHz
•
500 MHz Models: 500 MHz
•
350 MHz Models: 350 MHz
•
200 MHz Models: 200 MHz
•
100 MHz Models: 100 MHz
•
70 MHz Models: 70 MHz
10
Referencing the frequency from step 9, set the power
meter Cal Factor % to the frequency value on the
calibration chart on the power sensor.
11
Set the oscilloscope sweep speed according to the
following:
•
1 GHz Models: 500 ps/div
•
500 MHz Models: 1 ns/div
•
350 MHz Models: 2 ns/div
•
200 MHz Models: 2 ns/div
•
100 MHz Models: 5 ns/div
•
70 MHz Models: 5 ns/div
12
Note the oscilloscope Std Dev(1) reading at the bottom of
the screen.
13
Note the reading on the power meter and covert to Vrms
using the expression:
14
Calculate the response using the expression:
Example
If:
Pmeas
1_MHz
= 892 uW
Std Dev(n)
1MHz
= 210.4 mV
Vin
maxfreq
Pmeas
maxfreq
50
Ω
×
=
response(dB) = 20 log
10
MHz
1
M Hz
1
freq
max
freq
max
Vin
/
Vout
Vin
/
Vout
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...