2000/3000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
5
Check the voltages on the line filter board’s interboard
supply connector J108.
Figure 17
Line filter board interboard supply connector J108
a
If the VP13V bulk supply was bad at the system board
test point, check it again at the interboard supply
connector J108:
VP13V
VP1V8
VP1V4
VP1V
VP12V
VP1V2
VP1V
GND
LINE_TRIG
VP5V
VP3V3
VP2V5
39-29-9202
1252-8073
J108
2
5
10
7
4
8
19
16
13
3
18
1
15
9
6
11
14
17
20
12
Cable-Interboard Header
Test Point
(near J3701)
Expected Value
Assemblies
Supplied
If Good
If Bad
VP13V
+13 V ±3%
Whole
system
If good when the interboard supply
cable is disconnected from the
system board, but bad when the
interboard supply cable is
connected, this indicates a
problem with either the cable or
the assemblies supplied,
1 First, test the interboard supply
cable for opens or shorts using
the DMM.
2 If the interboard supply cable is
good, replace the system board.
Go to
“To check the line filter
board AC output”
on page 93 to
test whether the AC input to the
power supply is good.
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...