2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
e
Adjust the oscilloscope’s trigger level so that the rising
edge of the waveform at the center of the screen is
located where the center horizontal and vertical grid
lines cross (center screen).
f
Ensure the horizontal position control is set to
0.0 seconds.
4
Make the measurement.
a
Set oscilloscope horizontal sweep speed control to
1 ms/div.
b
Set horizontal position control to +1 ms (rotate control
CCW).
c
Set the oscilloscope horizontal sweep speed control to
5 ns/div.
d
Record the number of nanoseconds from where the
rising edge crosses the center horizontal grid line to
the center vertical grid line. The number of
nanoseconds is equivalent to the time base error in
ppm.
e
Use the date code on the oscilloscope’s serial tag to
calculate the number of years since manufacture.
Include any fractional portion of a year.
f
Use the following formula to calculate the test limits.
Time base accuracy limit: 25 ppm ±5 ppm per year
(aging)
g
Record the result and compare it to the limits in the
Performance Test Record (see
page 63
).
Date Code:
0747
07 = Year
47 = Week
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...