48
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div,
and 5 mV/div, noise may be a factor. To eliminate the noise,
add a BNC Tee, blocking capacitor, and shorting cap at the
oscilloscope channel input to shunt the noise to ground. See
Figure 6
. If a BNC capacitor is not available, use an SMA
blocking capacitor, adapter, and cap. See
“Blocking capacitor
and shorting cap”
in the equipment list on
page 26
for
details.
Figure 6
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor