132
2000/3000 X-Series Oscilloscopes Service Guide
5
Replacing Assemblies
To retreive the 40-pin display cable
1
Once the 75019- 68712 display assembly has been
removed, use a TORX T10 remove the two screws that
secure the display to its mounting bracket.
Figure 55
Disassembling the 75019-68712 display assembly
2
Carefully separate the upper and lower halves of the
display mounting bracket.