2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Figure 2
Setting Up Equipment for Digital Channel Threshold Accuracy
Test
4
Use a BNC- banana cable to connect the multimeter to the
other side of the BNC Tee.
5
Connect the BNC Tee to the Channel 1 output of the
calibrator as shown in
Figure 2
.
6
On the oscilloscope, press the
[Digital]
key, then press the
Thresholds
softkey, then press the
D7 - D0
softkey repeatedly
until the check mark is next to
User
.
7
Press the
User
softkey to the right of the
D7 - D0
softkey,
then turn the Entry knob (
) on the front panel of the
Probe
Cables
Channels
8 - 15
Channels
0 - 7
Test
Connector
2000/3000 X-Series Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC-Bananna
Cable