Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
29
To test digital channels (MSO models only)
The acquisition system testing provides confidence that the
acquisition system is functioning correctly. It does not,
however, check a particular specification.
1
Disconnect all probes from the circuit under test and from
any other input source.
2
Using probe leads and grabbers, connect digital channels
D0, D1, D2, and D3 to the Probe Comp signal on the
center of the front panel.
3
Press the
[AutoScale]
key.
If four square waves appear, the acquisition system is
functioning correctly.
If the square waves do not appear, go to the
“Troubleshooting” chapter. Then return here to finish
testing the digital channels.
4
Disconnect the digital channels from the calibration point.
5
Use steps 2 and 3 to test the following sets of digital
channels. After you test one set of digital channels,
remove them before connecting the next set.
•
D4, D5, D6, D7
•
D8, D9, D10, D11 (on 3000 X- Series models)
•
D12, D13, D14, D15 (on 3000 X- Series models)
Содержание 2000 X
Страница 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Страница 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Страница 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Страница 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Страница 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...